Description
The Electrolyte Analyzer YR06407 // YR06417 is an advanced device designed to provide precise and reliable results in measuring electrolytes in various clinical samples. This device features an interactive human-machine menu with dynamic real-time visualization of sample identifiers. It also includes automatic liquid level detection and alarms, along with real-time diagnostics of the system status. One of the standout features is its ability to detect and filter small bubbles, preventing blockages and ensuring accurate measurements. Its wave theory rinsing method and direct pipe rinsing method help prevent blockages and cross-contamination, maintaining the integrity of samples.
The YR06407 // YR06417 performs automatic calibrations and two-point corrections, offering printouts of quality control charts and statistical parameters for efficient monitoring. The power failure protection prevents the loss of critical data, and the device supports fuzzy queries, LIS software, and selectable data upload formats. Additionally, it manages the RTC clock and allows shutdown at any time to reduce reagent consumption, making it suitable for various hospital settings. This analyzer is compatible with serum, plasma, whole blood, and diluted urine samples, and can store up to 10,000 test results.
Market Price
The price of the Electrolyte Analyzer YR06407 // YR06417 varies depending on the supplier and additional specifications required, generally ranging from $2,600 to $7,000 USD. This price range reflects the advanced technology used, high-performance capabilities, and additional features included in the device. To get an exact and personalized quote, we invite you to generate an automatic quote through our Kalstein Plus platform, where you can access competitive prices and financing options tailored to your needs.
Frequently Asked Questions
- What types of samples can the YR06407 // YR06417 analyze?
The YR06407 // YR06417 is compatible with serum, plasma, whole blood, and diluted urine samples.
- How many results can the YR06407 // YR06417 store?
The analyzer can store up to 10,000 test results in its memory.
- What is the measurement speed of the YR06407 // YR06417?
The device can perform measurements in less than 30 seconds.
Advantages and Disadvantages
Advantages:
- High Precision: Automatic liquid level detection and bubble filtering.
- Efficiency: Automatic calibration and two-point correction.
- Flexibility: Compatible with multiple types of samples and analysis methods.
- Safety: Power failure protection and fuzzy queries.
Disadvantages:
- Initial Cost: The investment can be high for some small laboratories.
- Dependency on Specific Technology: Requires consumables and regular calibrations.
Field Use
The Electrolyte Analyzer YR06407 // YR06417 is widely used in clinical and hospital laboratories for precise electrolyte measurement in various samples. Its ability to handle a high volume of samples and provide quick results makes it ideal for high-demand environments. The integration of automatic calibration, bubble detection, and power failure protection functions ensures an efficient and reliable workflow, enhancing the laboratory’s productivity and precision.
Recommendations
To maximize the performance of the Electrolyte Analyzer YR06407 // YR06417, it is recommended to perform regular maintenance and follow the manufacturer’s usage guidelines. Ensure that the operating environment meets the specified temperature and humidity conditions (5-40°C, <80% RH) to guarantee accurate results. Additionally, train staff in the proper use of the instrument to take full advantage of its features and maintain the quality of analyses.
For more information and to request a personalized quote, visit our Kalstein Plus platform and get details on prices and financing options. Don’t miss the opportunity to improve your laboratory’s efficiency and precision with the YR06407 // YR06417.
Technical Specifications
Model | YR06407 | YR06408 | YR06409 | YR06410 | YR06411 |
Sample | serum, plasma, whole blood and dilute urine | ||||
Storage | up to 10000 test results | ||||
Printer | Internal thermal printer | ||||
Interface | RS232 port | ||||
Analysis Method | ion selective electrode (ISE) | ||||
Measuring speed | <30s | ||||
Sample position | 39 positions (including 5 emergency and 4 functional positions) | ||||
Temperature | 5-40°C | ||||
Relative Humidity | <80% | ||||
Atmospheric pressure | 86~106 kPa | ||||
Power supply | AC220V±22V, 50Hz±1Hz | ||||
Power | <120W | ||||
Test Parameters | K Na Cl | K Na Cl TCO AG | K Na Cl iCa nCa TCa pH | K Na Cl iCa nCa TCa pH TCO AG | K Na Cl Li |
Model | YR06412 | YR06413 | YR06414 | YR06415 | YR06416 | YR06417 |
Sample | serum, plasma, whole blood and dilute urine | |||||
Storage | up to 10000 test results | |||||
Printer | Internal thermal printer | |||||
Interface | RS232 port | |||||
Analysis Method | ion selective electrode (ISE) | |||||
Measuring speed | <30s | |||||
Sample position | 39 positions (including 5 emergency and 4 functional positions) | |||||
Temperature | 5-40°C | |||||
Relative Humidity | <80% | |||||
Atmospheric pressure | 86~106 kPa | |||||
Power supply | AC220V±22V, 50Hz±1Hz | |||||
Power | <120W | |||||
Test Parameters | K Na Cl iCa nCa TCa pH Li | K Na Cl iCa nCa TCa pH Li TCO AG | K Na Cl iCa nCa TCa pH Mg | K Na Cl iCa nCa TCa pH Mg TCO AG | K Na Cl iCa nCa TCa pH Mg Li TCO AG |
K Na Cl iCa nCa TCa pH Mg Li
|
Accessories
Model | Description |
YRA264 | K Electrode |
YRA265 | K Electrode |
YRA266 | Cl Electrode |
YRA267 | Ca Electrode |
YRA268 | pH Electrode |
YRA269 | Li Electrode |
YRA270 | Mg Electrode |
YRA271 | Ref. Electrode |
YRA272 | Reagent Pack Calibration A 430ml; Calibration B 110ml; ( For Model H900 only) |
YRA273 | Reference Electrode Internal Solution – 10ml |
YRA274 | Electrode Internal
Solution – 3ml |
YRA275 | Activating Solution – 110ml |
YRA276 | QC Solution Level 2
(Water-based) – 110ml |
YRA277 | Cleaning Solution – 110ml |
YRA278 | Electrode Wash Solution (Protein Enzyme) -125mg (25mg/bottle x5) |
YRA278-1 | Electrode Wash Solution (Diluent) – 5ml ( 1 bottle) |